Signal

 
This control tab offers settings for the signal to be measured.
 
Signal
1

Channel

 
1. Channel
The InnoAnalyzer 3D is a one-channel analysis tool. If more channels shall be viewed simultanously, the instrument can to be cloned.
By right-clicking on the channel, its color in the diagram can be adjusted (Change color).
2

Signal input

 
2. Signal input
Signal source selects the physical input for the vibration signal. Gain is used for the manual range selection.
3

Measurand and unit

 
3. Measurand and unit
 
The connected sensor determines the available options. Velocity and displacement are calculated from acceleration by integration (InnoAnalyzer 3D Pro only). The unit should be selected in a way that the measuring value does not drop below 0.001 permanently.
4

Characteristic

 
4. Characteristic
 
The following characteristics are available:
Peak value:
The hight of the FFT line represents the amplitude of the corresponding sinusoidal signal.
Peak-to-peak value:
The hight of the FFT line represents the peak-to-peak amplitude of the corresponding sinusoidal signal.
True r.m.s. value:
The hight of the FFT line represents the true r.m.s. value of the corresponding sinusoidal signal.